ナビゲーション


BL36XU OUTLINE

問い合わせ番号

INS-0000001555

Outline of Beamline

 Beamline BL36XU was constructed by the University of Electro-Communications and Institute for Molecular Science with support of RIKEN and JASRI under a New Energy and Industrial Technology Development Organization (NEDO) program, and moved to RIKEN beamline at March, 2020. BL36XU provides time-resolved QXAFS using tapered-Undulator beam and channel-cut monochromators, full-field/scanning XAFS imaging, XES, simultaneous time-resolved QXAFS/XRD, and pink beam experiment under in-situ experimental conditions.

Research Target

  • Time-resolved QXAFS
  • Full-field/scanning XAFS imaging
  • XES (HERFD-XANES, RIXS)
  • Simultaneous time-resolved QXAFS/XRD
  • Pink beam experiment

Key words

  • Measurement methods
    XAFS, In-situ, operando, time-resolved XAFS, 100 nm focused beam, 3-dimensional XAFS-CT imaging, XES, HERFD-XANES, RIXS, XRD, pink beam
  • Equipment
    transmission/fluorescence mode XAFS measurement equipment, XAFS-imaging system, XES measurement equipment, reaction gas supply and removal equipment, cryostat, high-temperature gas cell, fuel cell power generation equipment, polymer electrolyte fuel cell.

Source and Optics

  • Source
    In-vacuum type tapered undulator
  • Main optics
    ・Main optics are 1st mirror (M1), 2nd mirror (M2), two compact monochromators, high-energy-resolution monochromator for HAPES, 3rd mirror (M3) and 4th mirror (M4).
    ・The M1 and compact monochromators are cooled with liquid nitrogen. M2, M3 and M4 are cooled with water.
    ・All mirrors are coated with stripes of Pt and Rh. The M1 is a horizontal deflection flat mirror, and M2 is a spherically bent flat mirror for horizontal focusing. The M1 and M2 are placed apart at a distance of 7.5 m with a glancing angle of 2 mrad to separate reflected X-rays by 30 mm in the horizontal direction from high energy X-rays and gamma-rays emitted from the light source. The M3 and M4 are spherically bent flat mirrors for vertical focusing and higher harmonics rejection.
    ・The compact monochromators are driven with a servomotor. Two monochromators having a channel-cut Si (111) crystal and a Si (220) crystal are tandemly arranged to cover an energy range from 4.5 to 35 keV.

Layout of beamline

Optics hutch

Experimental hutch

Experimental station

  • QXAFS with time resolution of 20 ms is available using channel-cut crystal monochromators.
  • XAFS imaging with spatial resolution of 100 nm is available for the energy range of 4.5 to 35 keV using two KB mirror systems.
  • Full-field XAFS imaging with spatial resolution of 1 µm/100 nm is available.
  • Available X-ray detectors are fast ionization chambers, 25-element Ge detector, 4-element silicon drift detector, 2-dimensional pixel array detector, indirect X-ray imaging detectors.
  • Sample environment controlling equipment are a cryostat (4 K-RT), reaction gas supply and removal equipment, high-temperature gas cell (RT-1000 K), fuel cell and power generation equipment.

Experimental station

PUBLICATION SEARCH

* Sorry, Some parts of results are displayed using Japanese characters.

BL36XU PUBLICATION SEARCH

CONTACT INFORMATION

Please note that each e-mail address is followed by "@spring8.or.jp."

Tomoya URUGA
SPring-8 / JASRI
1-1-1 Kouto, Sayo-cho, Sayo-gun, Hyogo 679-5148
Phone : +81-(0)791-58-0833
Fax : +81-(0)791-58-0830
e-mail : urugat

Kiyofumi NITTA
SPring-8 / JASRI
1-1-1 Kouto, Sayo-cho, Sayo-gun, Hyogo 679-5198
Phone : +81-(0)791-58-0833
Fax : +81-(0)791-58-0830
e-mail : nittak

Kotaro HIGASHI
SPring-8 / JASRI
1-1-1 Kouto, Sayo-cho, Sayo-gun, Hyogo 679-5198
Phone : +81-(0)791-58-0833
Fax : +81-(0)791-58-0830
e-mail : higashi

最終変更日